4-Step Calibration Block
Model: 4-Step
Unit: Inch
Step Thickness: 0.250in, 0.500in, 0.750in, 1.000in
Usage: Ultrasonic calibration block step style are used to calibrate and verify the operation of ultrasonic thickness and flaw gauges.
Model: 4-Step
Unit: Inch
Step Thickness: 0.250in, 0.500in, 0.750in, 1.000in
Usage: Ultrasonic calibration block step style are used to calibrate and verify the operation of ultrasonic thickness and flaw gauges.




YUSHI offers a comprehensive range of step calibration blocks in various thickness steps and materials, along with custom manufacturing services. All the Blocks in accordance with ASTM E797-05.
Should you not find the required type in our catalog, please contact me at any time through the contact information on the website.
4-Step Calibration Block

5-Step Calibration Block

6-Step Calibration Block
![]() | ![]() |
| 6-Step 1mm, 3mm, 5mm, 10mm, 15mm, 20mm | 6-Step 2.5mm, 5mm, 10mm, 15mm, 20mm, 25mm |
7-Step Calibration Block
![]() | ![]() |
| 7-Step 1mm, 1.5mm, 2mm, 4mm, 6mm, 8mm, 10mm | 7-Step 3mm, 12mm, 24mm, 30mm, 36mm, 42mm, 48mm |
8-Step Calibration Block
![]() | ![]() |
| 8-Step 1mm, 2mm, 3mm, 4mm, 5mm, 6mm, 7mm, 8mm | Ultra Thin 8-Step 0.030'', 0.040'', 0.050'', 0.060'', 0.070'', 0.080'', 0.090'', 0.100'' |
10-Step Calibration Block

![]() | ![]() |
| VersaTip Calibration Block | Curved Step Calibration Block |
ABS Block Case
![]() | ![]() |
| Block Case for 4-Step, 5-Step and 6-Step | Block Case for 7-Step, 8-Step and 10-Step |
Core Applications of Ultrasonic Thickness Step Wedges
Thickness Gauge Calibration
Establish time-to-thickness correlation for material velocity settings
Verify measurement accuracy per ASTM E797 standards
Linearity Validation
Check instrument linearity across thickness ranges
Identify dead zones and measurement blind spots
Transducer Performance Testing
Evaluate probe resolution
Validate near-surface measurement capability
Reference Standardization
Provide traceable thickness references
Support ISO 17025-compliant calibration certificates
Material-Specific Calibration
Compensate for acoustic velocity variations
Thickness step wedges provide essential graduated reference standards for calibrating ultrasonic thickness gauges, validating measurement linearity, compensating for material velocity differences, and ensuring compliance with international inspection codes - particularly where ±0.1mm measurement accuracy is critical.
Add:U39, No. 81-43, Puhe Road, Hushitai Street, Shenbei New District, Shenyang, Liaoning, China 110136
E-mail:athena.w@yushindt.com
WeChat ID:965317198
Mobile / Whatsapp: +86 18602468800