【New Product】PR20 Ultrasonic Pulser Receiver with Adjustable Multi-Stage Transmitter Voltage
YUSHI has launched a new product - - Ultrasonic Pulser Receiver PR20.
This upgraded model features 8 adjustable transmitter voltage levels and an optimized design compared to the PR10. Below are the detailed specifications of the PR20.

The PR20, developed by YUSHI INSTRUMENTS, is a high-performance computer-controlled ultrasonic pulser/receiver. It includes a Windows-based control software for configuring trigger sources, transmitter voltage, repetition frequency, pulse width, single/dual crystal modes, damping, and gain adjustments to meet diverse application needs. Users can also leverage the open SDK to develop custom software on mainstream Windows systems.
1. Applications
Connecting an ultrasonic probe and an oscilloscope can help perform ultrasonic flaw detection, thickness measurement, probe detection, etc. It has a wide range of uses in non-destructive testing, material analysis and other fields.
2. Parameter Introduction
2.1 Transmitter Parameters
The transmitting part is used to adjust the various transmitting parameters of the ultrasonic pulse.
Transmitter Voltage
8 adjustable levels: 60V, 110V, 150V, 200V, 250V, 300V, 350V, 400V.
350V and 400V are unavailable when damping is set to 34Ω.
400V is unavailable when damping is 50Ω/90Ω.
Pulse Width
Adjustable from 1× to 40× (1× ≈ 28ns).
Limited to 1×–10× when damping is 510Ω and voltage is 400V.
Limited to 1×–10× when damping is 50Ω/90Ω and voltage is 350V.
Repetition Frequency
Eight options: 4Hz, 8Hz, 16Hz, 20Hz, 100Hz, 500Hz, 1kHz, 2kHz.
Trigger Source
Supports internal trigger (self-generated pulses) or external trigger (requires external signal generator).
2.2 Receiver Parameters
Controls the ultrasonic pulse receiver functionality.
Single/Dual Crystal
Single Crystal (Pulse-Echo Mode): Pulse transmission and echo reception via T/R interface.
Dual Crystal (Pitch-Catch Mode): Pulse transmission via T interface and reception via R interface.
Damping
4 options: 34Ω, 50Ω, 90Ω, 510Ω.
Gain
Adjustable from 0 to 60dB with sliding control or 1dB step increments.
3. Functional Applications
3.1 Probe Testing
Principle:
By performing a Fast Fourier Transform (FFT) on the ultrasonic echo signal, the system generates a frequency spectrum to determine the center frequency, which is then compared to the probe’s nominal frequency for verification.
3.2 Thickness Measurement
Principle:
Utilizes ultrasonic pulse reflection to measure waveform travel time, enabling calculation of material thickness.
4. Technical Specifications
4.1 Control Parameters
| Section | Parameter | Range | Step |
| Transmitter | Voltage | 60V, 110V, 150V, 200V, 250V, 300V, 350V, 400V | Selection |
| Pulse Width | 1×–40× (1× ≈28ns) | 1× | |
| Repetition Frequency (Internal) | 4Hz, 8Hz, 16Hz, 20Hz, 100Hz, 500Hz, 1kHz, 2kHz | Selection | |
| Trigger Source | Internal/External | Selection | |
Receiver | Single/Dual Crystal | Single (1T/1R), Dual (1T/2R) | Selection |
| Damping | 34Ω, 50Ω, 90Ω, 510Ω | Selection | |
| Gain | 0–60dB | 1dB |
4.2 Technical Parameters
| Pulse Type | Negative square wave |
| Rise Time | 2.0–7.5ns (≥20ns at 510Ω damping) |
| Fall Time | ~2.8ns |
| Pulse Width | 1× ≈28ns |
| Bandwidth | 0.5–20MHz |
| Excitation Mode | Pulse-Echo (P-E), Through-Transmission (T-T) |
| Repetition Frequency | Internal: 4Hz–2kHz; External: 0–20kHz |
| External Trigger | Rising edge trigger; Trigger voltage: 3.3V |
Notes:
Higher transmitter voltage increases rise/fall times.
Wider pulse width reduces transmitter voltage.
Higher repetition frequency slightly reduces transmitter voltage.
4.3 Specifications
| Dimension | 18.4cm × 12.6cm × 5cm | Operating Temp. | 0–55℃ |
| Weight | 542g |
4.4 Standard Accessories
| Item | Quantity |
| PR20 | 1 |
| BNC - BNC Cable | 2 |
| Communication USB Drive | 1 |
| USB Type-C Data Cable | 1 |
| Control Software (on USB) | 1 |